Light microscopy

From Libre Pathology
Revision as of 15:55, 27 May 2011 by Michael (talk | contribs) (→‎References: +EL)
Jump to navigation Jump to search

This article examines light microscopy.

Resolution

 .[1]
Where:

  •   = resolving distance; smaller better.
  •   = numerical aperture of the objective; typically 0.25 - 1.4, >1.0 is oil immersion, it is usu. inscribed on the lens itself.
  •   = numerical aperture of the condenser.
  •   = wave length of light.

It follows from the above equation that, closure of the condenser diaphragm results in a loss of resolution, i.e. R is larger.[1]

Stated differently:[2][3]

  • Opening the condenser --> increases resolution & brightness -- but -- decreases depth of field (DOF) & contrast.
  • Closing the condenser --> increases DOF & contrast -- but -- decreases resolution & brightness.

Numerical aperture

NA = numerical aperture.

General formula for NA:[4]
 .

Where:

  • n = index of refraction, n = 1.0 for air.
  • theta = half-angle of the max. cone of light

NA and f-number

N = f/D.

Where:

  • N = f-number, e.g. f 1.2, f 1.4, f 11.
    • Smaller N = larger opening.
  • f = focal length.
  • D = diameter of entrance pupil.

At infinity:
 .
 .
 .

Numerical aperture

If one substitutes the above into the equation at the top:
 .

Notes:

  • Larger 'D' is better.
  • Larger NA = better.

Lenses

  • Most lens = 'achromats' -- only correct green.
  • 'Apochromatic' lenses - correct all colours; very expensive.

Condenser

  • Condenser -- large flattened lens beneath the specimen.
    • Iris diaphragm.
      • Condenser diaphragm --> incr. contrast for resolution ---- large dia. good resol. bad contrast?
        • Field aperature diaphragm --> optical illumination.

Depth of field

  • Abbreviated DOF.
  • It depends on the aperature (small is better).[3]
    • Inverse relationship with resolution and brightness.
    • Related to contrast.

 .[5]

Kohler illumination

Rationale

  • Maximize resolution. (???)

Procedure

  1. Any specimen on stage.
  2. Focus.
  3. Adjust field aperture (bottom) - to obscure periphery of field of view (FOV).
  4. Raise or lower condenser until field aperture diaphragm clearly focused.
  5. +/-Center 'field aperture diaphragm - using condenser centering screws.

Resolution

  • Usual light microscopes are limited to about 0.2 micrometres.
    • Coming is "Super-resolution microscopy" - using high speed CCDs (charge-coupled devices).[6]